Evaluation of new technique of sterilization using biological indicator
Nomal Chintan Sheth, Yogesh V Rathod, Pratima R Shenoi, Deepa D Shori, Rajiv T Khode, Amruta P Khadse
Department of Conservative Dentisty and Endodontics, VSPM Dental College and Research Center, Nagpur, Maharashtra, India
Correspondence Address:
Dr. Nomal Chintan Sheth B2 Madhu Kunj, Navroji Cross Lane, Ghatkopar (w), Mumbai - 400 086, Maharashtra India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/JCD.JCD_253_16
|
|
Background: A novel technique of sterilization of endodontic files is introduced in this article.
Aims: Newly introduced sterilization unit, named “SteriFast” is compared with autoclave and glass bead sterilizer using biological indicator.
Materials and Methods: Spore strips of Bacillus pumilus were cultured in nutrient broth. This cultured media was used to contaminate the experimental samples of endodontic files. These contaminated files were sterilized using three different techniques. The sterilized files were transferred into nutrient medium under aseptic condition. The results were observed after 24 h, 48 h, and 7 days.
Results: The results showed that autoclave and new sterilization device (SteriFast) showed complete sterilization. The files sterilized using glass bead sterilizer showed bacterial growth (80%).
Conclusions: Thus, it proves that autoclave and SteriFast are ideal techniques of sterilization of endodontic files. Glass bead sterilizer does not completely sterilize the files. The article also compares SteriFast and autoclave in other aspects such as its design, basic principle, advantages, and disadvantages. The article also describes features and design of SteriFast, used for all kind of small dental instruments. |
|
|
|