Comparative evaluation of surface hardness and depth of cure of silorane and methacrylate-based posterior composite resins: An in vitro study


Department of Conservative Dentistry and Endodontics, Sharad Pawar Dental College, Sawangi, Wardha, Mumbai, Maharashtra, India

Click here for correspondence address and email

Date of Submission14-Dec-2014
Date of Decision18-Feb-2015
Date of Acceptance19-Feb-2015
Date of Web Publication12-Mar-2015
 

   Abstract 

Aim: This in vitro study was carried out to compare the effect of LED light curing system on polymerization and hardness of silorane-based and methacrylate-based posterior composite resin.
Materials and Methods: A total of 40 samples, 20 of silorane-based composite Filtek P-90 and 20 of methacrylate-based composite Heliomolar HB measuring 2 mm thickness and 8 mm diameter were prepared using Teflon molds and cured using LED curing light. The samples were polished and tested in Knoop hardness tester using a 50-gram load and dwell time of 15 seconds on top and bottom surfaces. The percentage depth of cure was calculated, and statistical analysis was performed using two-way ANOVA test and Student t- test.
Results: Higher statistically significant values were seen for both the top and bottom surface hardness in silorane-based resins than methacrylate-based resins on LED light curing.
Conclusion: Greater depth of cure was achieved in silorane-based posterior composite than in methacrylate-based posterior composite resins with a statistically significant difference.

Keywords: Depth of cure; methacrylate-based posterior composites; posterior composites; Silorane-based posterior composites; surface hardness

How to cite this article:
Agrawal A, Manwar NU, Hegde SG, Chandak M, Ikhar A, Patel A. Comparative evaluation of surface hardness and depth of cure of silorane and methacrylate-based posterior composite resins: An in vitro study. J Conserv Dent 2015;18:136-9

How to cite this URL:
Agrawal A, Manwar NU, Hegde SG, Chandak M, Ikhar A, Patel A. Comparative evaluation of surface hardness and depth of cure of silorane and methacrylate-based posterior composite resins: An in vitro study. J Conserv Dent [serial online] 2015 [cited 2023 Dec 4];18:136-9. Available from: https://www.jcd.org.in/text.asp?2015/18/2/136/153070

   Introduction Top


The search for an ideal restoration led to the introduction of composite resins by Bowen in 1962 and its continuous evolution since then to this present day. A variety of composites, including conventional, microfilled, hybrid, flowable, packable, nanofilled have been introduced till date. Though they have become the material of choice for anterior restorations, they were found lacking in strength to suit the requirements of a posterior restorative material. Various modifications in chemical structure and composition have been tried to overcome this issue. [1]

With the intention of extending the use of microfilled composite for posterior restoration along with anterior restorations, the proportional composition of the monomer mixture, i.e., the portion of comparatively large copolymers was lowered while the portion of microfillers was slightly raised by manufacturers. Also, a rheology modifier in the form of an organically modified compound silicate was added to the material. [1]

Currently, most composite resins have their common basis in the radical polymerization of methacrylates. Mozner and Salz (2001) [2] have reported that vinyl cyclopropane derivatives are also able to copolymerize with common methacrylate-based resins. The reaction involves the cationic ring opening of spiro ortho carbonates especially in combination with epoxy monomers. Silorane (3M ESPE), a low shrinkage tooth-colored restorative material, as claimed by the manufacturer, has been recently introduced in the market. Silorane was so named to indicate a hybrid compound of siloxane and oxirane functional moieties. Studies in the past have shown this material to be superior to conventional methacrylate-based composites. [3] Dye penetration tests have proved silorane-based composites to show significantly less microleakage at both the resin-enamel and resin-cementum interface. [4]

Solid state light emitting diodes (LED) have been used for curing composites since 1995. LEDs emit light at a specific wavelength of 470 nm required for activating camphoroquinone. These lamps offer high-energy performance, have a longer life span of more than 10,000 hours and undergo little degradation over time. Also, they consume little power and can be battery-operated. [5]

The degree of cure of resin composites influences its mechanical properties, solubility, dimensional stability, color change and biocompatibility. [6] Depth of cure and microhardness testing have been reliably and widely used to assess the relative degree of cure of resins, and thus the efficiency of light sources. The Knoop microhardness test has been shown to be one of the best indirect methods for testing the hardness of resin composite. [7]

The present in vitro study was carried out to compare the surface hardness and depth of cure achieved in methacrylate-based and silorane posterior composite resin on curing with LED curing unit.


   Materials and methods Top


[Table 1] shows all the equipment and materials tested in the study. A total of 40 samples were prepared. Teflon moulds with inner diameter of 8 mm and thickness of 2 mm were prepared. Moulds were placed on Mylar strips placed on a glass slab and were filled with composite using bulk insertion technique. In all, 20 samples were prepared using silorane-based Filtek P-90 posterior composite A3 shade (3M ESPE St. Paul, MN, USA) and 20 samples were prepared using methacrylate-based Heliomolar HB posterior composite, A3 shade (Ivoclar Vivadent AG, Schaan, Germany). A Mylar strip was placed on the upper surface and the material was flattened with a glass slide. After the removal of the excess material the glass slide was removed, leaving the Mylar strip. LED curing light (Rotex, India), with a light intensity of 1000 mw/cm 2 and a tip diameter of 8 mm was used to cure the samples for a period of 40 seconds with the tip kept at a distance of 1 mm. The intensity of the light curing unit was checked by an external hand-held radiometer before curing each specimen.
Table 1: Descriptive statistics and comparative analysis of top and bottom surface hardness values by Bonferroni test

Click here to view


The specimens were then removed from the molds, and polished and measured using digital Vernier calipers. The top surface was marked with an indelible marker. All samples were stored dry at room temperature in light-proof containers for 24 hours.

Knoop hardness testing was performed using a 50-gram load and a dwell time of 15 seconds. For each sample, three indentations were made on both the top and the bottom surfaces. The mean hardness values for the top and bottom surfaces were noted and percentage depth of cure was calculated using the following formula:



The data was statistically analyzed using a two-way ANOVA test, Bonferroni test and Student t-test. All the data collected was tabulated and statistically analyzed using SPSS (Statistical Package for Social Science) 14.0 software.


   Observations and result Top


[Table 1] shows the mean top and bottom surface hardness values between the groups. Results showed greater surface hardness for the silorane-based group (Group I) than the methacrylate-based group (Group II) with a p value of 0.000 (highly statistically significant) on both the top and bottom surfaces.

[Table 2] shows the comparison of top and bottom values in each group and the comparison of depth of cure. The difference in top and bottom surface hardness was statistically significant in both the groups. However, the difference in mean depth of cure was not statistically significant between silorane- and methacrylate-based groups.
Table 2: Descriptive statistics and comparative analysis of depth of cure by Bonferroni test

Click here to view



   Discussion Top


Siloranes, a new class of ring-opening monomers, were synthesized to overcome the problems related to polymerization shrinkage. Obtained from the reaction of oxirane and siloxane molecules, the volumetric shrinkage of a silorane-based composite was determined to be 0.99 vol. %. This resin is considered to have combined the two key advantages of the individual components: Low polymerization shrinkage due to oxirane monomers and increased hydrophobicity due to the presence of siloxane species in its composition. [3]

Apart from the predominant radical polymerization initiation in conventional methacrylate resin-based composites (RBCs), the silorane composite polymerizes by a cationic process, which is insensitive to oxygen. The silorane-based composite revealed decreased water sorption, solubility and associated diffusion coefficient compared with conventional methacrylate RBCs. [3] Nanjundasetty et al., [4] have shown superior marginal integrity both below and above the CEJ with siloranes compared to methacrylate-based composites.

The present study showed that greater hardness and depth of cure values were achieved in silorane-based posterior composite samples than with methacrylate-based posterior composite samples when cured with LED light.

Customized Teflon moulds were used, as the material does not stick to the mold and can be easily demolded after polymerization. As bulk insertion technique was adopted in this study and a maximum of 2 mm is acceptable, the thickness of the samples was standardized to 2 mm. [8]

The International Organization for Standardization (ISO) has standardized the specifications for light curing units to have a minimum intensity of 300 mW/cm 2 in 400-515 nm wavelength bandwidth at the curing light tip. [9] These criteria set by the ISO were fulfilled by the LED curing unit used in this study.

Light cured composites polymerize both during and after visible light activation, known as light and dark reactions, respectively. In all composites, maximum hardness is attained after 24 hours. [10] The samples were stored in dark light-proof boxes after curing in this study in order to prevent further curing due to exposure to light after the stipulated curing cycle.

The hardness of the composite is directly related to the degree of polymerization, and thus a good indicator of the degree of conversion of composite resins and a valuable parameter to estimate the mechanical properties. DeWald and Ferracane [11] have stated that Knoop hardness correlates well with the degree of conversion. Also, it minimizes the effect of elastic recovery, is a relatively simple technique and shows reliability of obtained result, hence it was the method chosen in this study.

The results of this study are in agreement with those of Guiraldo et al. [12] Composite hardness is influenced by several factors, such as organic matrix composition, type and amount of filler particles, and degree of conversion. The organic matrix of Filtek P-90 is composed mainly of silorane resin and inorganic particles, quartz and yttrium fluoride. In contrast, the organic matrix of Heliomolar is composed mainly of Bis-GMA, UDMA and 1,10- decandiol dimethacrylate, and a combination of inorganic particles silicon dioxide, ytterbium trifluoride and copolymers. Heliomolar HB is a microfilled composite. Several studies have suggested that microfilled composites are more difficult to cure because the small filler particles cause light to scatter, thus decreasing the effectiveness of the curing light. On the other hand, silorane-based composites have cationic polymerization reaction, which is characterized by continuous ring opening expansion initiated at the time of curing and promoted further cross-linking and hardening of the entire resin matrix.


   Conclusion Top


Within limitations of the study, higher top and bottom hardness values were achieved in silorane-based posterior composites as compared to methacrylate-based posterior composites, though both attained similar percentage depth of cure.

 
   References Top

1.
Roberson TM, Heymann HO, Ritter AV. Introduction to composite restorations. Sturdevant's Art and Science of Operative Dentistry. 5 th ed. St. Louis: Mosby; 2006. p. 495-526.  Back to cited text no. 1
    
2.
Mozner N, Salz U. New developments of polymeric dental composites. Prog Polym Sci 2001;26:535-76.  Back to cited text no. 2
    
3.
Al-Boni R, Raja OM. Microleakage evaluation of silorane based composite versus methacrylate based composite. J Conserv Dent 2010;13:152-5.  Back to cited text no. 3
[PUBMED]  Medknow Journal  
4.
Nanjundasetty JK, Nanda S, Panuganti V, Marigowda JC. Marginal sealing ability of silorane and methacrylate resin composites in class II cavities: A scanning electron microscopic study. J Conserv Dent 2013;16:503-8.  Back to cited text no. 4
[PUBMED]  Medknow Journal  
5.
Malhotra N, Mala K. Light curing considerations for resin- based composite materials: A review Part I. Compend Contin Educ Dent 2010;31:498-505.  Back to cited text no. 5
    
6.
Ceballos L, Fuentes MV, Tafalla H, Martinez A, Flores J, Rodriguez J. Curing effectiveness of resin composites at different exposure times using LED and halogen units. Med Oral Patol Oral Cir Bucal 2009;14: 51-6.  Back to cited text no. 6
    
7.
Price RB, Fahey J, Felix CM. Knoop microhardness mapping used to compare the efficiency of LED, QTH and PAC curing lights. Oper Dent 2010;35:58-68.  Back to cited text no. 7
    
8.
Yap AU. Effectiveness of polymerization in composite restoratives claiming bulk placement: Impact of cavity depth and exposure time. Oper Dent 2000;25:113-20.  Back to cited text no. 8
    
9.
Fan PL, Schumacher RM, Azzolin K, Geary R, Eichmiller FC. Curing light intensity and depth of cure of resin based composites tested according to international standards. J Am Dent Assoc 2002;133: 429-34.  Back to cited text no. 9
    
10.
Albers HF. Resin polymerization. Tooth colored restoratives. 9 th ed. London: BC Decker Inc.; 2002. p. 81-111.  Back to cited text no. 10
    
11.
DeWald JP, Ferracane JL. A comparison of four modes of evaluating depth of cure of light activated composites. J Dent Res 1987;66: 727-30.  Back to cited text no. 11
    
12.
Guiraldo RD, Consani S, Consani RL, Berger SB, Mendes WB, Sinhoreti MA, et al. Comparison of silorane and methacrylate based composite resins on the curing light transmission. Braz Dent J 2010;21:538-42.  Back to cited text no. 12
    

Top
Correspondence Address:
Shubha G Hegde
Department of Conservative Dentistry and Endodontics, Sharad Pawar Dental College, Sawangi, Wardha - 442 004, Mumbai, Maharashtra
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-0707.153070

Rights and Permissions



 
 
    Tables

  [Table 1], [Table 2]

This article has been cited by
1 Synthesis of Bis-GMA Grafted Co-Polymer of Acrylic–Itaconic Acid and its Composite
Mayuri Gupta, A. K. Tyagi, Manoj Raula
Polymer Science, Series B. 2022;
[Pubmed] | [DOI]
2 Effect of the Sample Preparation and Light-curing Unit on the Microhardness and Degree of Conversion of Bulk-fill Resin-based Composite Restorations
SSL Braga, ACT Schettini, ELO Carvalho, CAK Shimokawa, RB Price, CJ Soares
Operative Dentistry. 2022; 47(2): 163
[Pubmed] | [DOI]
3 Influence Of Different Led Curing-Units On Depth Of Cure and Micro-Hardness Of Nano-Hybrid Resin Composite
Alper KAPDAN, Seher KAYA, Digdem EREN
Cumhuriyet Dental Journal. 2021;
[Pubmed] | [DOI]
4 Effects of alternatively used thermal treatments on the mechanical and fracture behavior of dental resin composites with varying filler content
Ricardo Carrillo-Cotto, Adriana F. da Silva, Cristina P. Isolan, Raquel P.G. Selayaran, Maicon Selayaran, Fábio G. Lima, Eliseu A. Münchow
Journal of the Mechanical Behavior of Biomedical Materials. 2021; 117: 104424
[Pubmed] | [DOI]
5 Shear bond strength of three solvent-containing adhesive with different bulk fill composites
Gökhan Dokumacigil, Pinar Yilmaz Atali, Faik Bülent Topbasi
Journal of Adhesion Science and Technology. 2021; 35(5): 522
[Pubmed] | [DOI]
6 Effect of preheating and shade on surface microhardness of silorane-based composites
Marios Theodoridis, Dimitrios Dionysopoulos, Eugenia Koliniotou-Koumpia, Pavlos Dionysopoulos, Paris Gerasimou
Journal of Investigative and Clinical Dentistry. 2017; 8(2)
[Pubmed] | [DOI]



 

Top
 
 
 
  Search
 
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


    Abstract
   Introduction
    Materials and me...
    Observations and...
   Discussion
   Conclusion
    References
    Article Tables

 Article Access Statistics
    Viewed3375    
    Printed130    
    Emailed0    
    PDF Downloaded204    
    Comments [Add]    
    Cited by others 6    

Recommend this journal