1 Department of Conservative Dentistry and Endodontics, Gitam Dental College and Hospital, Visakhapatnam, Andhra Pradesh, India 2 Department of Conservative Dentistry and Endodontics, Vishnu Dental College, Bhimavaram, Andhra Pradesh, India 3 Department of Periodontics, Institute of Dental Sciences, Bhubaneshwar, Odisha, India
Correspondence Address:
C H Swarupa Door No. 34-13-38, Rossilloon Street, Gnanapuram, Visakhapatnam, Andhra Pradesh India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0972-0707.117506
|
|
Objective: To compare the efficacy of an integrated apex locator and an apex locator and endodontic motor assembly in maintaining the working length when operated under autoreverse mode.
Study Design: Thirty distobuccal roots of intact maxillary first molars were taken and access cavities were prepared. The teeth were divided into Group I: Prepared with TCM Endo V and Group II: Prepared with ProPex and NSK assembly. The instrumentation was ended in ProTaper F3 file, which was cemented in the canal. The roots were sectioned, observed under a stereomicroscope and the distance from instrument tip to the apical foramen was measured.
Results: Mean difference in the deviation of two groups was 0.075 mm, P = 0.34 (>0.05) which was statistically insignificant when assessed with unpaired t-test.
Conclusion: The assembly of ProPex-NSK Endo-mate DT and the apex locating endomotor TCM Endo V Nouvag are clinically acceptable. |
|
|
|