ORIGINAL ARTICLE
Year : 2013  |  Volume : 16  |  Issue : 6  |  Page : 555-558

An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators


1 Department of Conservative Dentistry and Endodontics, Institute of Dental Studies and Technologies, Modinagar, Uttar Pradesh, India
2 Consultant Endodontist, New Delhi, India
3 Department of Conservative Dentistry and Endodontics, ITS Dental College, Muradnagar, Uttar Pradesh, India
4 Department of Periodontics, Institute of Dental Studies and Technologies, Modinagar, Uttar Pradesh, India

Correspondence Address:
Nikhil Puri
Department of Conservative Dentistry and Endodontics, Institute of Dental Studies and Technologies, Modinagar, Ghaziabad - 201 204, Uttar Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-0707.120953

Rights and Permissions

Aim: The aim of the present study is to evaluate and to compare the accuracy of two electronic apex locators (EALs): DentaPort ZX and iPex, at a position 0.5 mm short of the apical foramen. Materials and Methods: Thirty single-rooted, mandibular premolar teeth were selected. Standard access cavities were prepared and the actual length (AL) was calculated. The samples were then embedded in alginate and the electronic measurements were determined and recorded. Results: The results obtained showed that in determining the root canal length with a tolerance level of ±0.5 mm, i.e., AL ± 0.5 mm, DentaPort ZX was accurate in 93.3% of the samples and iPex was accurate in 90% of the samples at a position 0.5 mm short of the apical foramen. Conclusion: A strong correlation was seen between the two electronic methods and AL and also in between the two EALs, showing the possibility of their use to measure the root canal length. No statistically significant difference was found between both the apex locators.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed1789    
    Printed32    
    Emailed0    
    PDF Downloaded171    
    Comments [Add]    

Recommend this journal